Avoiding jetting in mass flow controllers: the vena contracta

High flow mass flow controllers always need downstream valves and high flow meters can’t be plumbed with small process ports.   

The reason for this is jetting. But what is jetting? What causes it? Why is it important for Alicat devices in particular and why is it mainly an issue in high flow devices?  

To answer these questions, we need to first discuss what is happening as fluids pass through our valves. 

Categorizing flow through flow restrictions

In unrestricted laminar flow, fluid moves in a consistent and organized way. The flow channels run parallel in the direction of flow, such as in non-turbulent river and plumbing systems. The flow pattern is regular and easy to categorize. 

As flow approaches any restriction, such as an orifice or a valve, the outer flow channels, or those farthest away from the restriction, start to bend towards the inner flow channels, or those nearest to the restriction. As flow reaches to the restriction, the inner flow accelerates as a result of condensing flow from the outer flow channels. 

Just after the restriction, the outer flow channels continue to bend into the inner flow channels, resulting in continued narrowing and acceleration of the flow. Eventually, flow begins to move back from the inner flow channels to fill out the outer flow channels, returning to normal flow.  In other words, even though the flow path is again unrestricted and normal, the flow doesn’t instantly correct itself. Instead, it takes a little distance and time to spread out again.  

Vena contracta 

Due to fluid momentum, as fluid passes a restriction such as an orifice or valve, it continues to compress itself. Since this momentum takes some time and distance to change back around to start expanding, the fluid’s narrowest contraction point is just past the orifice or valve. 

We refer to this narrowest flow point as the vena contracta, which means “narrow veins” in Latin. The term jetting is used to describe the phenomena of this unique, constricted flow pattern.   

Adjusting for jetting in Alicat mass flow devices  

Now that we’ve explained jetting and the vena contracta, let’s discuss its implications for Alicat devices. 

In Alicat devices, jetting happens when high flow gases or liquids pass through Alicat valves, restricting the flow and taking some distance and time to refill the whole flow channel back up, since the vena contracta is located after the valve. If the flow remains constricted for too long and doesn’t expand to fill the entire flow channel inside Alicat devices, then flow rates become inaccurate. Because the geometry that the flow follows is different than what the Alicat device is expecting, the values which the Alicat device reports for flow rate become largely meaningless.  

To avoid problems which may occur because of jetting, Alicat devices at high flow rates use downstream valves and use larger orifices, reducing the distance that it takes for jetting to dissipate, ensuring that flow rate measurements are highly accurate since the jetting does not have an impact on the flow measurement.  

This is not usually an issue which occurs in lower flow devices. Since the vena contracta’s distance from the valve or orifice is dependent on flow rate, occurring closer to the valve or orifice at lower flow, at lower flow rates, flow can normalize primarily before reaching Alicat’s laminar flow elements. With high flow rates, the vena contracta moves further downstream of the flow restriction, taking more distance to expand all the way back out. For this same reason, small process ports can’t be used on high-flow devices, as there just isn’t enough time for everything to expand back out.  

More support 

Alicat’s application engineers are prepared to help you avoid jetting by picking the correct device configurations for your needs, ensuring consistent, accurate, and repeatable flow measurement and control.  

Contact an applications engineer for flow and pressure solutions today  